Strictly comply with ASTM standards for identifying, testing and evaluating the quality, characteristics and parameters of materials, products, systems or services
Basic Order Information
According to ASTM f2063-18 or GB 24627-2009 standard, basic information should be provided when making inquiry or placing order for wire or pipe
Certificate and Package Label
Certificate of conformity, product inspection report will be provided for each shipment
Applicable Standards and Documents
● ASTM F2063 Standard Specification for Wright Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants, or equivalent GB 24627-2009
● ASTM F2633 Standard Specification for Wrought Seamless Nickel-Titanium Shape Memory Alloy Tube for Medical Devices and Surgical Implants.
● ASTM F2005-05 Standard Terminology for Nickel-Titanium Shape memory Alloys
● ASTM F2004 Standard Test Method for Transformation Temperature of Nickel-Titanium Alloys by Thermal Analysis
● ASTM F2516 Standard Test Method for Tension Testing of Nickel-Titanium Superelastic Materials
● ASTM F2082/F2082M Standard Test Method for Determination of Transformation Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free Recovery
Ordering Information on Nitinol Wires and Tubes
Per ASTM F2063-12 or GB 24627-2009, ordering information or request for quotation should contain following info:
● Chemical composition or alloy code;
● Transformation temperature for the material at delivery status, such as Af
● Shape of the material such as bars, wires, tubing, or others
● Dimensions, such as diameter, OD & ID of tube, as well as their tolerances
● Heat Treatment status at delivery, such as Super-elastic, Cold Worked
● Surface status at delivery
● Mechanical Properties: When delivered in super-elastic state, upper plateau stress, tensile strength and elongation should be specified; when delivered in cold-worked state, tensile strength and elongation should be specified
● Quantity, such as weight, length, or number of pieces
Certification and Labeling Information
A certification indicating that the material meets the requirements of specification, plus test reports shall be given with shipment.
Testing Report: In terms of superelastic wires or tubes, following data should be given
● Dimension
● UTS
● Elongation
● Upper plateau stress
● Active Af by Bend & Free Recovery
● Permanent set after straining to 6% strain
● Chemical analysis
Labeling Information
● Product Code
● Product Name
● Lot #
● Raw Material
● Inspection Report #
● QTY such as meter, kilogram, piece